Reports
> Medical
> Abstract of "Anesthesia Alarms in Context..."
Title: Anesthesia Alarms in Context: An Observational Study Abstract This paper surveys current work on the design of alarms for anesthesia environments and notes some of the problems arising from the need to interpret alarms in context. Anesthetists' responses to audible alarms in the operating room were observed across four types of surgical procedure (laparoscopic, arthroscopic. cardiac, and intracranial) and across three phases of a procedure (induction, maintenance, and emergence). Alarms were classified as (a) requiring a corrective response, (b) being the intended result of a decision, (c) being ignored as a nuisance alarm, or (d) functioning as a reminder. Results revealed strong effects of the type of procedure and phase of procedure on the number and rate of audible alarms. Some alarms were relatively confined to specific phases; others were seen across phases and responses differed according to phase. These results were interpreted in light of their significance for the development of effective alarm systems. Actual or potential applications of this research include the design of alarm systems that are more informative and more sensitive to operative context than are current systems. To purchase a copy of this article, click the Buy Now link below. An electronic (PDF) copy will be sent within 24 hours of order receipt. To avoid delivery delays, refer to the file size indicated above and check your e-mail system message and storage capacity before ordering. (Note: Cost of this article includes a copyright fee paid to the original copyright owner). Changes last made October, 2007 Copyright © 2004, 2007 Beta Research, Inc.. All rights reserved. |